PANDAA Lassa Virus - Rotor-Gene Q ====================== For Rotor-Gene Q 5-Plex with 72-Well Rotor 1. Download the .ret template file from https://www.aldatubio.com/qpcr-templates/ RUN SETTINGS ====================== To prevent any run setting errors, avoid manually programming the Rotor-Gene Q and ensure that the template PANDAA Lassa Virus - RGQ Template.ret is used. The Auto-Gain prompt that appears at the start of a new run should be set so that auto-gain optimisation is performed before the 1st acquisition. Do NOT perform gain optimisation at 60 °C at the beginning of the run. ANALYSIS SETTINGS ====================== It is not possible to save analysis settings in the .ret template file, so when analysing the results: 1. Dynamic Tube should be selected. 2. Slope Correct should be selected for consistent results across all runs. 3. Outlier Removal should be set at 5%. Target Name: PANDAA Lassa Virus (LASV) Method: Dynamic Exclusion: Extensive Fluorescence Cutoff: 5% Threshold: Refer to PANDAA Lassa Virus instructions for use. Target Name: Internal Control (IC) Method: Dynamic Exclusion: None Fluorescence Cutoff: N/A Threshold: Auto Target Name: Sample Adequacy Control (SAC) Method: Dynamic Exclusion: None Fluorescence Cutoff: N/A Threshold: Auto